BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Catalog Number 381467
Device Problems Device Disinfection Or Sterilization Issue (2909); Device Packaging Compromised (2916)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/22/2018
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone #: (b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that the packaging of a bd insyte¿ autoguard¿ bc shielded iv catheter was found degraded and opening spontaneously breaching sterility.There was no report of exposure, injury or medical intervention.
 
Manufacturer Narrative
H.6.Investigation summary: device/batch history record review findings: review was performed on the lot number 5232532.The product was manufactured on afa line 6 from august 27, 2015 thru august 31, 2015 and packaged on packaging line 11 from september 1, 2015 thru september 2, 2015.Review of the dhr¿s revealed all required challenges samples and testing was performed per specification in accordance with the set-up and in process sampling plans.Per review it was noted that there were no reject activity findings throughout the build of this lot that would impact upon the quality of the product.Set-up and in-process inspections included but were not limited to blister thickness, bad seal/cut/holes, seal transfer width and package leak test.These were performed on various stages throughout the process, all the inspections passed per specifications.No significant issues were found during the review of the device history record.The peura (end user risk analysis) was analyzed to determine the risk to customer.The analysis showed that current risk is acceptable.Occurrence and severity rankings have not changed.Observations and testing: received one unused iag 14ga unit in partially opened package from the lot number; 5232532.Visual/microscopic examination: the package was partially opened at the top of the blister pack.The analysis of top web adhesive: the product characteristics require a minimum of 1/8¿ seal transfer.Per observation of the photos provided this characteristic was met.In addition, the bottom web in the photos revealed evidence of adequate transfer (adhesive) from the paper top web at the open seals and partially open seals of the packages (blister packs).The key variables that affect seal strength are: seal transfer/width and top web adhesive presence.Both of these variables were looked at during the investigation.Conclusions: the defect package seal integrity poor/questionable, stated in the description of the complaint was confirmed with the returned unit.Even though the package was partially opened, all the process characteristics that directly influence the seal were observed to be within specification.There was no physical evidence to confirm or to support manufacturing process related issues for the defect.Capa 48637 was initiated to address the issue of open seal complaints.Tasks associated with the capa plan have been completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE¿ AUTOGUARD¿ SHIELDED IV CATHETER
Type of Device
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
MDR Report Key7607643
MDR Text Key111845193
Report Number1710034-2018-00328
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier30382903814672
UDI-Public30382903814672
Combination Product (y/n)N
PMA/PMN Number
K952861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/31/2018
Device Catalogue Number381467
Device Lot Number5232532
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/04/2018
Initial Date Manufacturer Received 05/25/2018
Initial Date FDA Received06/15/2018
Supplement Dates Manufacturer Received05/25/2018
Supplement Dates FDA Received06/26/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-