• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. MAGEC SPINAL BRACING AND DISTRACTION SYSTEM; GROWING ROD SYSTEM-MAGNETIC ACTUATION Back to Search Results
Model Number MS1-5570R
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  Injury  
Event Description
It was reported that allegedly the patient's magec rod is damaged.The physician has revised the rod without incident.
 
Manufacturer Narrative
A visual inspection of the returned device revealed it was partially distracted with score marks.Functional testing revealed that the rod could lengthen by hand when pulling on the distraction rod, which confirms the pin failure.A device history record review revealed that the magec rod met all the required quality inspections and was released within specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SPINAL BRACING AND DISTRACTION SYSTEM
Type of Device
GROWING ROD SYSTEM-MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise, suite 100
aliso viejo CA 92656
MDR Report Key7607898
MDR Text Key111304764
Report Number3006179046-2018-00046
Device Sequence Number1
Product Code PGN
Combination Product (y/n)N
PMA/PMN Number
K140613
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberMS1-5570R
Device Catalogue NumberPA0564
Device Lot NumberA150220-03-00-01U
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/16/2018
Initial Date FDA Received06/15/2018
Supplement Dates Manufacturer Received05/16/2018
Supplement Dates FDA Received10/08/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Disability;
-
-