Catalog Number UNK HIP
Device Problem Defective Device (2588)
Patient Problems No Consequences Or Impact To Patient (2199); Not Applicable (3189)
Event Date 05/21/2018
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The reported hex-driver was used for tha surgery to insert the cup on (b)(6) 2018.It was reported that the screw driver tip could not detached from the screw head and the come out with the screw when insert the cup into the patient¿s hip joint.The surgeon tried to remove the screw by hand, but the driver and the screw head were still stick together.Finally, the surgeon pinched the screw with a pliers and hit it with a hammer to remove it.The surgery was finished without any other problem although it was not reported how the surgery was completed.There was less than 30min.Surgical delay and there was no adverse consequence to the patient.No further information was provided by the hospital.
 
Manufacturer Narrative
Additional narrative: product complaint #
=
> (b)(4).Investigation summary
=
> the device associated with this report was not returned.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP IMPLANT
Type of Device
HIP IMPLANT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46582-0988
5743725905
MDR Report Key7609402
MDR Text Key111838078
Report Number1818910-2018-62414
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 05/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK HIP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/21/2018
Initial Date FDA Received06/18/2018
Supplement Dates Manufacturer Received07/23/2018
Supplement Dates FDA Received07/27/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-