• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER LAUREATE WORLD PHACO SYSTEM; UNIT, PHACOFRAGMENTATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER LAUREATE WORLD PHACO SYSTEM; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750502
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Information (3190)
Event Date 05/21/2018
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
A customer reported that before surgery, the system froze.
 
Manufacturer Narrative
Additional information provided.The company service representative examined the system and was able to replicate the reported event.The host printed circuit board (pcb) was replaced.Unrelated to the reported event, the footswitch cable was replaced.The system was then tested and met all product specifications.The system was manufactured on december 16, 2014.Based on qa assessment, the product met specifications at the time of release.The root cause of the reported event was a nonconforming host printed circuit board (pcb).The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAUREATE WORLD PHACO SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
MDR Report Key7609771
MDR Text Key111845865
Report Number2028159-2018-01256
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
PMA/PMN Number
K021566
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Other
Type of Report Initial,Followup
Report Date 08/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065750502
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 05/20/2018
Initial Date FDA Received06/18/2018
Supplement Dates Manufacturer Received08/14/2018
Supplement Dates FDA Received08/30/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-