• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Catalog Number BI70002000
Device Problems Break (1069); Mechanical Problem (1384); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/13/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information not provided due to (b)(6) patient privacy regulations.The x-stage cover was returned to the manufacturer for analysis.Analysis found that the caster mount is physically damaged.The docking pin of left mount is broken.Analysis found that the reported event was related to a mechanical issue.After replacing the part, the imaging device performs as intended.
 
Event Description
Medtronic received information regarding an imaging device being used intra-operatively for a sacroiliac and thoracolumbar procedure.No surgical delays reported and no known impact on the patient outcome.It was reported that a nut on the image acquisition system (ias) dock was damaged.The x-stage cover was damaged and deformed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
stephanie riley
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key7610555
MDR Text Key111380765
Report Number1723170-2018-02546
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberBI70002000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/13/2017
Initial Date FDA Received06/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/10/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-