• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PALISADES DENTAL, LLC IMPACT AIR 45; HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PALISADES DENTAL, LLC IMPACT AIR 45; HANDPIECE Back to Search Results
Model Number 401
Device Problems Device Maintenance Issue (1379); Use of Device Problem (1670); Improper or Incorrect Procedure or Method (2017); Maintenance Does Not Comply To Manufacturers Recommendations (2974)
Patient Problem Foreign Body In Patient (2687)
Event Date 03/14/2018
Event Type  Injury  
Manufacturer Narrative
As part of palisades dental's investigation, on 01-jun-2018 a review of palisades dental sales records confirmed the sale of (b)(4) on 25-mar-2016 to (b)(4).In (b)(4).Records further indicate that the handpiece has not been at palisades dental for service since it was originally sold.A review of manufacturing records for (b)(4) did not find any abnormalities with the batch record or associated records.On 05-jun-2018 handpiece (b)(4) was received and evaluated.The following was found: cap and turbine are non-palisades dental parts.Palisades dental has determined that the failure of handpiece (b)(4) is not manufacturer related, rather it is related to service that did not comply to manufacturer's recommendations, and/or the use of non-palisades dental parts - any or all of these factors contributed to the alleged failure.Palisades dental has determined that either use of non-palisades dental parts and/or service that did not potentially contributed to the event.Palisades dental deems no action needed since the ifu instructs the user on returning the device to palisades dental for evaluation and service.
 
Event Description
It was reported that during an apicoetomy #3 a dentsply multi purpose surgical bur (mubur) came out of handpiece (b)(4) while doctor was using and the bur went into the sinus of a female patient.The bur was later removed by an oral surgeon and the patient is doing fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPACT AIR 45
Type of Device
HANDPIECE
Manufacturer (Section D)
PALISADES DENTAL, LLC
111 cedar lane
englewood NJ 07631
Manufacturer (Section G)
PALISADES DENTAL, LLC
111 cedar lane
englewood NJ 07631
Manufacturer Contact
maureen mcgovern
111 cedar lane
englewood, NJ 07631
2015690050
MDR Report Key7610695
MDR Text Key111382973
Report Number3003963943-2018-00003
Device Sequence Number1
Product Code EFB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K972376
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Dentist
Device Model Number401
Device Lot Number7289
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/01/2018
Initial Date FDA Received06/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/12/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-