MEDTRONIC MEXICO ACHIEVE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number 990063-020 |
Device Problem
Kinked (1339)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/28/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, a maneuvering issue was noted on the mapping catheter.Additionally, under fluoroscopy, the distal shaft of the mapping catheter appeared kinked.The catheter was removed and replaced.The case was completed with cryo.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Event summary: the patient data files showed at least nine applications were performed with catheter 2af284 / 81441 without any issue on the date of the event.The mapping catheter 990063-020/ 214483807 was not returned for investigation.In conclusion, investigation was not possible for the reported ¿achieve kink¿ issue.The mapping catheter was not returned for investigation.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|