• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS POWERGLIDE MIDLINE CATHETER FULL KIT; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS POWERGLIDE MIDLINE CATHETER FULL KIT; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number M118101
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/29/2018
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of an hub detachment is confirmed but the exact cause is unknown.Two photo samples of a powerglide catheter and extension set were returned for evaluation.An initial visual observation of the photos revealed the distal end of the extension tubing detached from the male luer connector.A ring of adhesive appears to be visible near the detached end.Based on the description of the reported event and photo samples provided, possible contributing factors include excessive tensile force and insufficient adhesive bonding.The complaint of a detached hub is confirmed but the exact cause could not be determined with the photo samples provided.A lot history review (lhr) of rebv0884 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported that the tubing of the extension set came apart from the hub connected to the powerglide catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERGLIDE MIDLINE CATHETER FULL KIT
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas
MX  
Manufacturer Contact
shelly gilbert
605 n. 5600 w.
salt lake city, UT 84116
8015225640
MDR Report Key7614778
MDR Text Key112001689
Report Number3006260740-2018-01437
Device Sequence Number1
Product Code FOZ
UDI-Device Identifier00801741038112
UDI-Public(01)00801741038112
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K133856
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM118101
Device Catalogue NumberM118101
Device Lot NumberREBV0884
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2018
Is the Reporter a Health Professional? Yes
Event Location Hospital
Initial Date Manufacturer Received 05/30/2018
Initial Date FDA Received06/19/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-