ZIMMER BIOMET, INC. DISC ULNA 3X155MM LT W/BRNG C; PROSTHESIS, ELBOW Back to Search Results
Model Number N/A
Device Problem Insufficient Information (3190)
Patient Problems Pain (1994); Loss of Range of Motion (2032); Sleep Dysfunction (2517)
Event Date 04/18/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: 211250, compr srs 50 mm dst hml bdy lt, 430570.Unknown, unknown comprehensive srs mod stem, unknown.Unknown, unknown comprehensive srs flange, unknown.Unknown, unknown discovery humeral condyle kit, unknown.(b)(6).Reported event was unable to be confirmed due to limited information received from the customer.The product was not returned for the investigation, therefore a visual and dimensional evaluations could not be performed.X-rays were submitted to mmi for review.Bone quality is osteopenic.Multiple images which demonstrate punctate hyperdensities within the soft tissues at the level of the elbow which could suggest metallosis.Dhr was reviewed and no discrepancies were found.Root cause could not be determined with information available.A possible contributing factor to the pain could be the metallosis.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.This report is being submitted late as it has been identified in remediation.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2018-01623, 0001825034-2018-04144, 0001825034-2018-04145, 0001825034-2018-04146.
 
Event Description
It was reported the patient reported moderate difficulty sleeping due to pain, being unable to do heavy chores, to carry a bag, to wash back, to cut, to do recreational activities, and extremely limited with daily activities.Patient also indicated during the 6 month follow up moderate arm, shoulder, or hand pain, and no pain at the one year follow up.Attempts have been made and no further information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISC ULNA 3X155MM LT W/BRNG C
Type of Device
PROSTHESIS, ELBOW
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7619078
MDR Text Key111655073
Report Number0001825034-2018-04143
Device Sequence Number1
Product Code KWT
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
PK043505
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,s
Reporter Occupation Physician
Type of Report Initial
Report Date 06/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date09/30/2017
Device Model NumberN/A
Device Catalogue Number114818
Device Lot Number869410
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/18/2016
Initial Date FDA Received06/20/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/15/2012
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age51 YR
Patient Weight80
-
-