• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problems Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/24/2018
Event Type  malfunction  
Manufacturer Narrative
Medtronic representative went to the site to test the system and reported that the site had been using fusion for more cases with no issues.Part not returned for evaluation.
 
Event Description
Medtronic received information regarding a navigation system during a functional endoscopic sinus surgery (fess) procedure.It was reported that the exams were already loaded and during registration, the system became unresponsive.After rebooting, the system seemed to function normally and about 3 hours into the case it became unresponsive again during navigation.The mouse was unresponsive as well.The surgery was completed successfully with the use of navigation and the delay was less than an hour.There was no known patient impact.
 
Manufacturer Narrative
Correction: a medtronic representative went to the site to test the equipment.Testing revealed that the reported issue could not be replicated.The system then passed the system checkout and was found to be fully functional.The logs for the navigation system were reviewed by medtronic personnel.However, the logs provided no additional insight into the probable cause of the anomaly.
 
Manufacturer Narrative
A software analysis was initiated to determine the probable cause of the issue through log analysis.Analysis found that the reported event was related to a software issue.This issue was documented in a medtronic navigation software anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7620254
MDR Text Key111702352
Report Number1723170-2018-02892
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 09/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/24/2018
Initial Date FDA Received06/20/2018
Supplement Dates Manufacturer Received09/05/2018
09/19/2018
Supplement Dates FDA Received09/10/2018
09/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/22/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age18 YR
Patient Weight45
-
-