It was reported that a patient underwent an atrial fibrillation (afib) procedure with a smartablate¿ irrigation tubing set, and when the tubing was connected to the saline and was flushed, bubbles were noticed and foreign material was found inside the tubing set.Loose and floating muddiness (suspended matter like dirt) was observed inside the tube.Flushing was performed before the tubing was used.No error code was observed on the irrigation pump.The device was replaced and the issue resolved with no patient consequence.The bubbles are not mdr reportable because bubbles in the tubing set can be an expected part of procedure set up, and flushing should be performed in order to remove the bubbles.However, the foreign material in the tubing is reportable.
|
It was reported that a patient underwent an atrial fibrillation (afib) procedure with a smartablate¿ irrigation tubing set, and when the tubing was connected to the saline and was flushed, bubbles were noticed and foreign material was found inside the tubing set.Additional information was received on 7/17/2018.The manufacturer date is 09/11/2017 and expiration date is 08/31/2018.The corresponding field of this report has been updated.The investigational analysis has been completed.The product was visually inspected and found to be in normal condition.No foreign material was found.Irrigation testing was performed and tubing passed specification.No micro bubbles were found in the tubing after flushing.The complaint was not confirmed.Manufacture ref no: (b)(4).
|