VYAIRE MEDICAL, INC 3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100
Device Problems Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/23/2018
Event Type  Injury  
Manufacturer Narrative
Vyaire medical reference number (b)(4).At this time, vyaire medical has not received the suspect device from the customer for evaluation.In the event that the device is received for evaluation or additional information is received, a follow-up report will be submitted.
 
Event Description
The customer reported that the piston kept stopping while the device was in use on a patient.The circuit lost pressure each time causing the piston to stop.There was no patient harm/injury associated with the reported issue.The customer reported calibrating the driver displacement indicator (ddi) board and completing the performance check.The customer reported that the device is working to specifications and cancelled the requested onsite field service.
 
Manufacturer Narrative
Results of investigation: this unit was field serviced by a vyaire medical authorized field service technician.The service technician calibrated the ddi board and performed performance checks on the ventilator, and the ventilator is working within specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATORY VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
avery foster
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key7621295
MDR Text Key111740085
Report Number2021710-2018-08003
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 06/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/23/2018
Initial Date FDA Received06/20/2018
Supplement Dates Manufacturer Received05/25/2018
Supplement Dates FDA Received03/08/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-