• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL, INC. SENSORMEDICS 3100A; VENTILATOR, HIGH FREQUENCY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL, INC. SENSORMEDICS 3100A; VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100A
Device Problems No Audible Alarm (1019); Disconnection (1171)
Patient Problem No Information (3190)
Event Date 06/08/2018
Event Type  malfunction  
Event Description
The map alarm was not audibly working.The map drifted, alarm light came on, but no audible alarm heard.Alarms set appropriately, but ventilator did not sound when alarms reached.Manufacturer response for ventilator, sensormedics 3100a (per site reporter).Unit under a service contract.Called for service.Biomed told to look inside for speaker wire disconnect.Wire was off.Reconnected the wire, plugged unit in then removed ac power.Got an audible power fail alarm which shows speaker now working.Unit returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSORMEDICS 3100A
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL, INC.
1100 bird center drive
palm springs CA 92262
MDR Report Key7628965
MDR Text Key112013962
Report Number7628965
Device Sequence Number1
Product Code LSZ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number3100A
Device Catalogue Number768901
Other Device ID Number08212/1152834
Was the Report Sent to FDA? Yes
Date Report Sent to FDA06/18/2018
Event Location Hospital
Date Report to Manufacturer06/18/2018
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
NO
-
-