• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. E1 RINGLOC BIPOLAR 28X51MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. E1 RINGLOC BIPOLAR 28X51MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Difficult to Insert (1316); Difficult To Position (1467)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that this bipolar liner did not insert into the bipolar shell.This surgery was finished with backup product.It was found that ring was bent.Attempts to obtain additional information have been made; however, no more is available.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Inspection of the returned device revealed there was a circular indentation on the outside radius of the liner.Device history record (dhr) was reviewed and no discrepancies were found.Root cause of the event could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E1 RINGLOC BIPOLAR 28X51MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key7629365
MDR Text Key112195385
Report Number0001825034-2018-04038
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
PMA/PMN Number
PN/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date06/07/2022
Device Model NumberN/A
Device Catalogue Number110010462
Device Lot Number266000
Other Device ID Number(01) 0 0880304 56878 5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2018
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/30/2018
Initial Date FDA Received06/22/2018
Supplement Dates Manufacturer Received10/10/2018
Supplement Dates FDA Received11/01/2018
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-