• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICMO13.2
Device Problem Inadequacy of Device Shape and/or Size (1583)
Patient Problems Intraocular Pressure, Delayed, Uncontrolled (1936); Pain (1994); No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4).
 
Event Description
The reporter indicated the surgeon implanted a 13.2mm vticmo13.2 implantable collamer lens, -15.5/+2.0/123 (sphere/cylinder/axis), in the patient's right eye (od), on (b)(6) 2017.The lens was explanted due to excessive vaulting and significant reduction of irido-corneal angles.The patient experienced ocular pain in darkness, with iop spikes.The lens was exchanged for a shorter lens and the problem was resolved.The reporter indicated the cause of the event was unknown.
 
Manufacturer Narrative
Device evaluation: lens was returned dry in a micro-centrifuge vial.Visual inspection found haptic torn and foreign material (residue) on lens surface.Claim # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
6263037902
MDR Report Key7629830
MDR Text Key112036710
Report Number2023826-2018-00945
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Expiration Date09/30/2020
Device Model NumberVTICMO13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/28/2018
Initial Date FDA Received06/22/2018
Supplement Dates Manufacturer Received07/02/2018
Supplement Dates FDA Received07/30/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTRIDGE MODEL SFC-45 - LOT # 1396843; FOAM TIP PLUNGER MODEL FTP - LOT # 1397838; INJECTOR MODEL AND LOT # UNK
Patient Outcome(s) Required Intervention;
-
-