Evaluation found an open slider in a patient access area which is not obvious to the user and could allow the unintentional patient escape.Additionally, five inches unraveling was found on both legs of the foot side panel.The stitching unraveling is in non-patient access areas and does not affect the functionality of the canopy.Posey beds are multi-use, serviceable items.As such, it is anticipated that the units may occasionally require repair, and as part of standard care the beds should be inspected prior to use.If damage is noted during these routine bed inspections, the unit should not be put into use with a patient and should be returned to posey for servicing.If the zipper slider body is bent open (i.E., the mouth is wider than manufacturing specifications) it can prevent the slider from properly engaging the zipper teeth.In areas where the slider engages the teeth, the zipper is secure and cannot be opened.However, if the teeth do not engage it can potentially leave an unsecured area.Applying pressure directly to the unsecured area will reveal the breach, which is why the user manual advises caregivers to apply direct pressure along the entire length of the zipper.Service issues are trended and reviewed by management on a monthly basis.As part of this monthly review, trends and excursions above control limits will be assessed, documented and acted upon as warranted.No corrective or preventative actions are necessary at this time.(b)(4).
|