Reportable based on analysis of the returned specimen.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.The history records indicate this product was final inspection tested and was determined to be acceptable.As received, the returned specimen consisted of one safari2 275cm sml crv; returned coiled, loose and double-bagged within "zip-lock" style poly biohazard pouches.The specimen presented bend damage in the large diameter curve and at 170.0 to 182.0cm from the distal aspect of the formed curve, and an offset overlapping coil condition located 186.25 to 186.40 cm from the distal aspect of the formed curve creating a localized oversized diameter of.04180".The ptfe coating in the vicinity of the offset/overlapping coil wraps is frayed and presents coating removal.No other damage or inconsistencies are noted to the specimen at this time.Both joints appear to be correct and intact by visual examination and by non-destructive testing.Except where noted, the specimen device appears visually and dimensionally correct.Based on the evidence presented by the sample and the information provided by the supporting documentation, clinical and/or procedural factors appear to have impacted on the event as reported.
|