Model Number 106
Device Problem Energy Output To Patient Tissue Incorrect (1209)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/25/2018
Event Type  malfunction  
Event Description
It was reported that, after a hysterectomy procedure on (b)(6) 2018, the patient suspected her vns device was not working.The patient's neurologist reported that the patient visited two days later ((b)(6) 2018) to check her vns device.The neurologist stated that upon interrogation it was found that the device's output current were had been set to 0ma due to an unknown reason.The exported programming data from the neurologist's programming tablet was received and reviewed.The data showed no indication that the device's output currents were disabled.There was no indication of a malfunction from the available data.No additional or relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key7631551
MDR Text Key112170129
Report Number1644487-2018-01069
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750061
UDI-Public05425025750061
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/27/2018
Device Model Number106
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 05/28/2018
Initial Date FDA Received06/22/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/10/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age37 YR
-
-