• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE CLINITRON RITEHITE C-8 ASSY; BED, AIR FLUIDIZED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE CLINITRON RITEHITE C-8 ASSY; BED, AIR FLUIDIZED Back to Search Results
Model Number P0800010019
Device Problem Unintended Movement (3026)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/26/2018
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the hand pendant needed to be replaced.Per the hill-rom service manual, hand pendant malfunction: 1.Replace the hand pendant with a known good one, and test each of its functions.(rga the pendant removed from the unit.) all functions operate normally with the new pendant.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the hand pendant to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating the head of bed went up by itself.The bed was located at the account.There was no patient/user injury reported.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITRON RITEHITE C-8 ASSY
Type of Device
BED, AIR FLUIDIZED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
erin padgett
1069 state route 46 east
batesville, IN 47006
8129312791
MDR Report Key7635125
MDR Text Key112315757
Report Number1824206-2018-00250
Device Sequence Number1
Product Code INX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K964223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 05/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberP0800010019
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/26/2018
Initial Date FDA Received06/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/29/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-