• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - SAN JOSE GALAXY INTRAVASCULAR ULTRASOUND SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - SAN JOSE GALAXY INTRAVASCULAR ULTRASOUND SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number H749A70200
Device Problem Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.If there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Event Description
Same case as mdr id#: 2134265-2018-06070 and 2134265-2018-06117 it was reported that automatic pullback failure occurred.An ilab ultrasound imaging system was used in conjunction with an opticross¿ imaging catheter and pullback sled to view the target lesion.The catheter was not recognized on the first case.After replacing the catheter, it was okay.However, on the second case, the catheter was not recognized and pullback failed.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAXY INTRAVASCULAR ULTRASOUND SYSTEM
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - SAN JOSE
150 baytech drive
san jose CA 95134
Manufacturer (Section G)
BOSTON SCIENTIFIC - SAN JOSE
150 baytech drive
san jose CA 95134
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7635509
MDR Text Key112295085
Report Number2134265-2018-06086
Device Sequence Number1
Product Code IYO
Combination Product (y/n)N
PMA/PMN Number
K980851
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 06/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH749A70200
Device Catalogue NumberA7020
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/06/2018
Initial Date FDA Received06/25/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-