Model Number 990063-020
Device Problem Kinked (1339)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/16/2018
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the device and data files were returned and analyzed.Data files showed 12 injections were performed with a balloon catheter on the date of the event without triggering of any system notice.Visual inspection of the mapping catheter showed loop kink.The mapping catheter was connected to the diagnostic computer and channel pairs did not display any noise.Movement and deflection of the catheter distal to the lemo connector did not induce any interference noise.In conclusion, the reported noise issue was not confirmed through data analysis or testing.The mapping catheter failed the return inspection due to a loop kink.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, noise was observed on the mapping catheter.The mapping catheter was replaced with resolve.The case was completed with cryo.No patient complications have been reported as a result of this event.The device was returned to the manufacturer, analyzed, and tested out of specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7635950
MDR Text Key112285827
Report Number9612164-2018-01535
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102588
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/16/2019
Device Model Number990063-020
Device Catalogue Number990063-020
Device Lot Number214246274
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/01/2018
Initial Date FDA Received06/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/16/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-