MEDTRONIC MEXICO ACHIEVE MAPPING CATHETER; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number 990063-020 |
Device Problem
Kinked (1339)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/16/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Product event summary: the device and data files were returned and analyzed.Data files showed 12 injections were performed with a balloon catheter on the date of the event without triggering of any system notice.Visual inspection of the mapping catheter showed loop kink.The mapping catheter was connected to the diagnostic computer and channel pairs did not display any noise.Movement and deflection of the catheter distal to the lemo connector did not induce any interference noise.In conclusion, the reported noise issue was not confirmed through data analysis or testing.The mapping catheter failed the return inspection due to a loop kink.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, noise was observed on the mapping catheter.The mapping catheter was replaced with resolve.The case was completed with cryo.No patient complications have been reported as a result of this event.The device was returned to the manufacturer, analyzed, and tested out of specification.
|
|