Model Number 990063-020
Device Problem Kinked (1339)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/31/2018
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the mapping catheter, 990063-020 with lot number 214988295, was returned and analyzed.Visual inspection of the mapping catheter showed the shaft and tip/ loop section were intact with no apparent issues.No kinks were observed along the shaft, or tip/ loop section of the mapping catheter.In conclusion, the kink was not confirmed through testing.The mapping catheter passed the returned product inspection as per specification.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, it was observed that the tip of the mapping catheter was kinked.The mapping catheter was replaced and the case was completed with cryo.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7637918
MDR Text Key112430408
Report Number9612164-2018-01540
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
K102588
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/06/2020
Device Model Number990063-020
Device Catalogue Number990063-020
Device Lot Number214988295
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/31/2018
Initial Date FDA Received06/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/06/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age67 YR
-
-