• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CASHEL 466P306X; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CASHEL 466P306X; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 466P306X
Device Problem Unintended Movement (3026)
Patient Problem Disability (2371)
Event Date 06/01/2018
Event Type  Injury  
Manufacturer Narrative
Additional information is pending and will be submitted within 30 days upon receipt.
 
Event Description
As reported by the legal department, the patient underwent placement of the trapease vena cava filter.The filter subsequently malfunctioned and caused injury and damages to the patient including, but not limited to, migration, unknown disability, and anxiety.As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, and pain and suffering, and other damages.
 
Manufacturer Narrative
The exact implant date is unknown.The catalog number is unknown, if received it will be provided.As reported, the patient underwent placement of a trapease vena cava filter.The filter subsequently malfunctioned and caused injury and damages to the patient including, but not limited to, migration, an unknown disability, and anxiety.As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, and pain and suffering, and other damages.The product was not returned for analysis.Additionally, as the sterile lot number was not available, device history record review could not be performed.The trapease filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.With limited information available it is unknown if the disability experienced is due to the filter.Clinical factors that may have influenced the event described include patient, pharmacological, or other comorbidities.The reported anxiety that the patient experienced does not represent a device malfunction.Inferior vena cava (ivc) filter migration is a known potential adverse event associated with all ivc filter implants and is listed in the instruction for use (ifu) as such.Possible causes for filter migration includes mega cava, wire entrapment during central venous catheter placement, ¿sail¿ effect (cranial migration) of large clot burden within the filter, mechanical device failure, and operator error.Physiologic causes of migration may result from temporary dysmorphism of the inferior vena cava including bending, coughing or valsalva maneuvers resulting in dislodgment of the filter.Some studies suggest that strenuous physical activity and increased intra-abdominal pressure can lead to migration of ivc filters.Without images or procedural films for review, the reported filter migration could not be confirmed, and the exact cause could not be determined.Giving the limited information available at this time, clinical factor contributing to the migration could not be determined.Also, with the limited information available for review at this time, there is nothing to suggest that the reported event is related to the design and/or manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
466P306X
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel, co. tipperary
Manufacturer (Section G)
CORDIS CASHEL
cahir road
cashel, co. tipperary
EI  
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014
7863138372
MDR Report Key7638128
MDR Text Key112335204
Report Number1016427-2018-01584
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number466P306X
Device Catalogue Number466P306X
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/01/2018
Initial Date FDA Received06/26/2018
Supplement Dates Manufacturer Received07/03/2018
Supplement Dates FDA Received07/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Disability;
-
-