It was reported that a patient underwent an ablation procedure for atrial fibrillation (afib) with a smartablate¿ irrigation tubing set and during the procedure, when the tubing set was primed, a powdery material was observed.The tubing set was replaced and the issue resolved.Tubing was never used on the patient and no patient consequences were reported.Multiple attempts have been made to gain clarification on this event with no response.There¿s no information on if the ¿powdery form¿ mentioned on the event description was inside or outside the tubing set.Therefore, this event will be conservatively reported as a foreign material found on the inside of the tubing.
|
It was reported that a patient underwent an ablation procedure for atrial fibrillation (afib) with a smartablate¿ irrigation tubing set and during the procedure, when the tubing set was primed, a powdery material was observed.Additional information was received on 7/17/2018, indicating the manufacture date of 10/02/2017 and the expiration date of 9/30/2018.Corresponding fields of this report have been updated.The device history record (dhr) was reviewed and no anomalies were found related to this complaint.In addition, the dhr review verifies that the device was manufactured in accordance with documented specification and procedures.The investigational analysis has been completed.The product was visually inspected and it was found to be in normal conditions.Flow testing was performed and the product passed all specifications.No error or bubble was found in tubing.Complaint was unable to duplicate.Manufacturer ref no: (b)(4).
|