RADIOMETER MEDICAL APS ABL800 FLEX ANALYZER; ABL800 FLEX ANALYZER WITH CREA MODULE Back to Search Results
Model Number 393-801
Device Problems Incorrect Or Inadequate Test Results (2456); Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/30/2018
Event Type  malfunction  
Manufacturer Narrative
Correction/removal reporting number is yet to be determined.Internal radiometer reference is (b)(4) and crr will be submitted no later than (b)(6) 2018.
 
Event Description
According to the complaint, a customer reports false creatinine results applying abl800 analyzer.From a comparison test, comparing creatinine results from cobas (roche) with those from abl800 analyzer demonstrate that creatinine concentration as determined on abl800 could be ~25% lower than on cobas.
 
Manufacturer Narrative
From a capa related to fan 915-387, it has been decided to increase lower limit for ac6+l1 (quality control) with 20 microm to reduce risk of low creatinine measurements.The final conclusion will be included in crr 915-387.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABL800 FLEX ANALYZER
Type of Device
ABL800 FLEX ANALYZER WITH CREA MODULE
Manufacturer (Section D)
RADIOMETER MEDICAL APS
aakandevej 21
broenshoej, 2700
DA  2700
MDR Report Key7641238
MDR Text Key112455050
Report Number3002807968-2018-00047
Device Sequence Number1
Product Code CHL
Combination Product (y/n)N
PMA/PMN Number
K041874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Recall
Type of Report Initial,Followup
Report Date 07/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number393-801
Device Catalogue Number393-801
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/30/2018
Initial Date FDA Received06/27/2018
Supplement Dates Manufacturer Received07/18/2018
Supplement Dates FDA Received07/30/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-