• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. POD6; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. POD6; HCG, KRD Back to Search Results
Catalog Number RBYPOD6
Device Problems Bent (1059); Physical Resistance (2578)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/31/2018
Event Type  malfunction  
Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.The hospital disposed of the device.
 
Event Description
The patient was undergoing a coil embolization procedure in the splenic artery using pod6s.During the procedure, the physician successfully deployed and detached the initials pod coils in the target vessel using a non-penumbra microcatheter.While advancing a pod6 through the microcatheter, the physician experienced resistance and upon pushing further, the pod6 pusher assembly became bent; therefore, it was removed.The procedure was successfully completed using additional pod coils and the same microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POD6
Type of Device
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key7642585
MDR Text Key112507622
Report Number3005168196-2018-01259
Device Sequence Number1
Product Code HCG
UDI-Device Identifier00814548013817
UDI-Public00814548013817
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Catalogue NumberRBYPOD6
Device Lot NumberF81238
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/31/2018
Initial Date FDA Received06/27/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/02/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age82 YR
-
-