• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Visual Prompts will not Clear (2281); Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/06/2018
Event Type  malfunction  
Manufacturer Narrative
The reported complaint of "system error, out of service, revert to manual cpr" error message was confirmed during the functional test and archive data review of the autopulse platform ((b)(4)).It was determined that the root cause was due to a defective processor board.Initial functional testing of the autopulse platform could not be performed due to the system error message on the screen that could not be cleared.Review of the system archive data revealed error messages (error 136 - internal parameter corrupted) multiple times that occurred on the event date.During visual inspection, the autopulse platform had no physical damage.The autopulse platform is a reusable device and was first manufactured on 24 sep 2004.The device has been operating for nearly 14 years and has exceeded its expected serviceable life of 5 years.Historical complaints were reviewed for service information related to the reported complaint and there was no previous history of complaint reported for autopulse platform with serial number (b)(4).
 
Event Description
As reported by a biomed, during shift check, the autopulse platform ((b)(4)) displayed "system error, out of service, revert to manual cpr" error message.Unable to clear error message by the user.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
kimthoa nguyen
2000 ringwood ave,
san jose, CA 95131
4084192922
MDR Report Key7644296
MDR Text Key112574366
Report Number3010617000-2018-00679
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000277
UDI-Public00849111000277
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 06/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-01-66
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/06/2018
Initial Date FDA Received06/27/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/24/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-