Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Abscess (1690); Erosion (1750); Unspecified Infection (1930); Hernia (2240); Blood Loss (2597)
Event Date 05/15/2016
Event Type  Injury  
Manufacturer Narrative
Unknown taper.The reporter of the event was asked to indicate product serial number.To date, the device information has not been received by apollo.Without the device serial number, the taper type is unknown.If returned, visual examination may determine the connector type associated with this event.This event was reported by the patient.To date, apollo has been unable to confirm the reported events with the patient's physician.Device labeling addresses the reported event as follows: precautions: over-dissection of the stomach during placement may result in slippage or erosion of the band and require reoperation.Anti-inflammatory agents, which may irritate the stomach, such as aspirin and non-steroidal anti-inflammatory drugs, should be used with caution.The use of such medications may be associated with an increased risk of erosion.Adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications which may result from the use of this product include the risks associated with the medications and methods utilized in the surgical procedure, the risks associated with any surgical procedure and the patient's degree of intolerance to any foreign object implanted in the body.There is a risk of band erosion into stomach tissue.Erosion of the band into stomach tissue has been associated with revision surgery, after the use of gastric irritating medications, after stomach damage and after extensive dissection or use of electrocautery, and during early experience.Symptoms of band erosion may include reduced weight loss, weight gain, access port infection, or abdominal pain.Re-operation to remove the device is required.Infection can occur in the immediate post-operative period or years after insertion of the device.In the presence of infection or contamination, removal of the device is indicated.Other adverse events considered related to the lap-band system that occurred in fewer than 1% of subjects included: esophagitis, gastritis, hiatal hernia, pancreatitis, abdominal pain, hernia, incisional infection, infection, redundant skin, dehydration, gi perforation, diarrhea, abnormal stools, constipation, flatulence, dyspepsia, eructation, cardiospasm, hematemesis, asthenia, fever, chest pain, incision pain, contact dermatitis, abnormal healing, edema, paresthesia, dysmenorrhea, hypochromic anemia, band leak, cholecystitis, esophageal dysmotility, esophageal ulcer, esophagitis, port displacement, port site pain, spleen injury and wound infection.Warnings: patients should be advised that the lap-band ap system is a long-term implant.Explant (removal) and replacement surgery may be indicated at any time.Medical management of adverse reactions may include explantation.Revision surgery for explantation and replacement may also be indicated to achieve patient satisfaction.
 
Event Description
Reported as: a patient with the lap-band system was reported to have a "band eroding in the stomach." the patient noted two months prior to removal they couldn't eat at all.Patient went to the er and an x-ray was performed.It noted the device was in place.A week later the patient went to see a physician who performed an endoscopy and it was observed the device had eroded into the stomach.The patient had the device removed.After removal, the patient went home and noted they were bleeding.Through blood work performed a week after the surgery, the bleeding was observed to be internal bleeding.The patient required a blood transfusion.The patient went to have a second surgery in which an abscess and six hernias were discovered.The patient also reported there were multiple infections through the body.At this time apollo has been unable to confirm the patient reported event with the surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAP-BAND SYSTEM
Type of Device
ADJUSTABLE GASTRIC BAND
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC
1120 s. capital of texas hwy
bldg. 1, ste. 300
austin TX 78746
Manufacturer (Section G)
ALLERGAN
global park free zone
900 global park
la aurora de heredia,
CS  
Manufacturer Contact
laura lebouef
1120 s. capital of texas hwy
bldg. 1, ste. 300
austin, TX 78746
5122795141
MDR Report Key7647872
MDR Text Key112702122
Report Number3006722112-2018-00155
Device Sequence Number1
Product Code LTI
Combination Product (y/n)N
PMA/PMN Number
P000008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial
Report Date 05/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 05/31/2018
Initial Date FDA Received06/28/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age66 YR
Patient Weight73
-
-