MAUDE Adverse Event Report: MALEM MEDICAL LIMITED MALEM; ALARM CONDITIONED RESPONSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MALEM MEDICAL LIMITED MALEM; ALARM CONDITIONED RESPONSE Back to Search Results
Device Problems Leak/Splash (1354); Overheating of Device (1437); Device Inoperable (1663)
Patient Problems Reaction (2414); Burn, Thermal (2530); Chemical Exposure (2570)
Event Date 06/23/2018
Event Type  Injury  
Event Description
Parents came in to the er with child who had suffered burns from the normal use of an enuresis alarm.It appears that the alarm somehow got very hot to a point where the plastic housing began to bend inwards from heat and caused the 2 housed batteries to leak onto the child's neckline.Child has suffered allergic reaction to the alarm and skin rash as a severe side effect of the same.Child was administered with first aid and discharged as per routine procedure.The alarm was returned to the parents.Device appears to have malfunctioned and overheated to hurt child in the neck.Missing on product.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MALEM
Type of Device
ALARM CONDITIONED RESPONSE
Manufacturer (Section D)
MALEM MEDICAL LIMITED
lowdham, nottingham NG14 7EJ
UK  NG14 7EJ
MDR Report Key7650120
MDR Text Key112851784
Report NumberMW5078139
Device Sequence Number1
Product Code KPN
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Nurse
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/28/2018
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age6 YR
Patient Weight22
-
-