• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ROCHE DIAGNOSTICS COBAS ELECSYS ANTI-TPO; SYSTEM, TEST, THYROID AUTOANTIBODY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ROCHE DIAGNOSTICS COBAS ELECSYS ANTI-TPO; SYSTEM, TEST, THYROID AUTOANTIBODY Back to Search Results
Catalog Number ASKU
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/23/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This event occurred in (b)(6).
 
Event Description
The customer received questionable results for multiple assays for one patient from cobas 8000 - cobas e 602 module serial number (b)(4).Refer to the attachment to the medwatch for all patient data.The medwatchs involved in this event include patient identifiers: (b)(6).The erroneous results were reported outside of the laboratory.There was no allegation of an adverse event.
 
Manufacturer Narrative
Sample from the patients was submitted for investigation.An interfering factor was found to be present in the sample.This interference is documented in product labeling for the assay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHE DIAGNOSTICS COBAS ELECSYS ANTI-TPO
Type of Device
SYSTEM, TEST, THYROID AUTOANTIBODY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key7650430
MDR Text Key113251904
Report Number1823260-2018-02134
Device Sequence Number1
Product Code JZO
Combination Product (y/n)N
PMA/PMN Number
K051890
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 07/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberASKU
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/12/2018
Initial Date FDA Received06/29/2018
Supplement Dates Manufacturer Received06/12/2018
Supplement Dates FDA Received07/17/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ADALAT OROS
Patient Age85 YR
-
-