• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL REVEAL LINQ; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number LNQ11
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Under-Sensing (1661); Invalid Sensing (2293)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/15/2018
Event Type  malfunction  
Manufacturer Narrative
This device was reported as included in the field action.Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action.It is included in the field action in the abundance of caution.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the implantable cardiac monitor (icm) experienced probable intermittent undersensing of r-waves.It was further reported the icm detected false tachycardia episodes due to oversensing of artifact/noise.The icm remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVEAL LINQ
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH   1131
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7651316
MDR Text Key112764517
Report Number9614453-2018-02448
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00643169476998
UDI-Public00643169476998
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K132649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial
Report Date 06/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/28/2017
Device Model NumberLNQ11
Device Catalogue NumberLNQ11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/19/2018
Initial Date FDA Received06/29/2018
Date Device Manufactured01/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1249-2016
Patient Sequence Number1
Patient Age87 YR
-
-