• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL CIRCULATION AUTOPULSE® RESUSCITATION MODEL 100; CARDIAC CHEST COMPRESSOR Back to Search Results
Model Number MODEL 100
Device Problems Visual Prompts will not Clear (2281); Device Displays Incorrect Message (2591); Device Operational Issue (2914)
Patient Problem No Patient Involvement (2645)
Event Date 06/05/2018
Event Type  malfunction  
Manufacturer Narrative
The customer's reported complaint of the autopulse platform displayed user advisory (ua) 41 (patient temperature sensor failure) error message was confirmed in the archive review and during the initial functional testing.Temperature sensor was replaced to remedy the fault.Visual inspection was performed and found damaged front and bottom enclosures, unrelated to the reported complaint.Front and bottom enclosures were replaced to address the issue.The autopulse platform is a reusable device and was manufactured in mar 2013.It has exceeded its expected service life of 5 years.Therefore, this type of physical damage found during visual inspection is characteristic of normal wear and tear for the life of the device and are unrelated to the reported complaint.During functional testing, the platform displayed (ua) 41 upon powering up, thus confirmed the reported complaint.Review of the archive data show (ua) 41 error messages occurred on the reported event date.After replacing the defective parts, the platform was further tested for 15 minutes with large resuscitation testing fixture, (lrtf) equivalent to 250 pound patient and passed all functional testing criteria and met all required specifications.Historical complaints were reviewed for service information related to the reported complaint and there was one similar complaint reported for autopulse with serial number (b)(4).Ccr (b)(4), reported on (b)(6) 2014.Temperature sensor was replaced to address this complaint.
 
Event Description
During shift check, autopulse platform ((b)(4)) was displaying an error message user advisory (ua) 41 (patient temperature sensor failure).Customer tried to clear the error, however, the issue was not resolved.No patient involvement on this issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION
2000 ringwood ave.
san jose CA 95131
Manufacturer Contact
noemi schambach
2000 ringwood ave,
san jose, CA 95131
4084192955
MDR Report Key7652994
MDR Text Key112891627
Report Number3010617000-2018-00696
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000512
UDI-Public00849111000512
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 06/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/05/2018
Initial Date FDA Received06/29/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/04/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-