U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNOVIS SURGICAL INNOVATIONS VASCULAR PROBE; DILATOR, VESSEL, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNOVIS SURGICAL INNOVATIONS VASCULAR PROBE; DILATOR, VESSEL, SURGICAL Back to Search Results
Catalog Number 7451015ES
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 03/06/2018
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that particulate matter (pm) was observed within the inner pouch of a vascular probe.The pm was observed during incoming inspection and before patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.Visual inspection revealed particulate matter on the outside wall of the inner pouch.Microscopic inspection revealed that the particulate matter was a fiber.The size of the particulate matter was found to be out of specification for this product.The reported particulate matter condition was verified.The cause of the condition was determined to be a manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASCULAR PROBE
Type of Device
DILATOR, VESSEL, SURGICAL
Manufacturer (Section D)
SYNOVIS SURGICAL INNOVATIONS
saint paul MN
Manufacturer (Section G)
SYNOVIS SURGICAL INNOVATIONS
2575 university ave w
saint paul MN 55114
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7655211
MDR Text Key113009880
Report Number1416980-2018-04003
Device Sequence Number1
Product Code DWP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K130896
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number7451015ES
Device Lot NumberSP18A021264951
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/21/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/04/2018
Initial Date FDA Received07/02/2018
Supplement Dates Manufacturer Received07/26/2018
Supplement Dates FDA Received08/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/02/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-