MAUDE Adverse Event Report: EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT WITH TRUWAVE DISPOSABLE PRESSURE TRANSDUCER AND VAMP JR.; TRANSDUCER, PRESSURE, CATHETER TIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES DR PRESSURE MONITORING KIT WITH TRUWAVE DISPOSABLE PRESSURE TRANSDUCER AND VAMP JR.; TRANSDUCER, PRESSURE, CATHETER TIP Back to Search Results
Model Number VMP306PX
Device Problem Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/15/2018
Event Type  malfunction  
Manufacturer Narrative
One single dpt - vamp jr.Kit was returned for examination.The reported event of damage to the device was confirmed.The pressure tubing was found broken from the bond joint connected to the lower tube of the proximal sample site.The cross surfaces of the broken tubing appeared uneven.In addition, the tubing had detached from the bond joint with a vamp reservoir stopcock.Indications of bonding solvent were evident on some locations of tubing bond surface area.The tubing outer diameter was measured near the point of detachment and was found to be within specification.A review of the manufacturing records indicated that the product met specifications upon release.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.It is common clinical practice to inspect all products before usage.These products are used by highly trained clinicians, experienced in identifying and mitigating any hazards that arise during use.If the pressure tubing becomes detached during use, it will affect the pressure waveform, which will immediately alert the clinician to begin the troubleshooting process.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised to consider the potential benefits in relation to the possible complications.In this event, there was no patient compromise noted as the issue was found before use.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.(b)(4).
 
Event Description
It was reported that upon opening a disposable pressure transducer vamp jr combination kit, a connector was observed to be ¿snapped¿ off.Another kit was used and worked successfully.No patient complications were reported as the issue was discovered before use.
 
Manufacturer Narrative
The investigation of this vamp jr.System leakage concluded that a potential root cause could be related to an incorrect solvent bonding execution during the assembly process.Personnel acknowledgement was conducted at the manufacturing site to prevent recurrence of this type of complaint.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRESSURE MONITORING KIT WITH TRUWAVE DISPOSABLE PRESSURE TRANSDUCER AND VAMP JR.
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
parque industrial de itabo
carr. sanchez km 18.5
haina, san cristobal
DR 
MDR Report Key7658734
MDR Text Key113027360
Report Number2015691-2018-02591
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
PMA/PMN Number
K925638
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup,Followup
Report Date 06/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/06/2019
Device Model NumberVMP306PX
Device Catalogue NumberVMP306PX
Device Lot Number61042025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2018
Initial Date Manufacturer Received 06/19/2018
Initial Date FDA Received07/03/2018
Supplement Dates Manufacturer Received07/18/2018
07/23/2020
Supplement Dates FDA Received08/16/2018
12/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-