It was reported that the patient¿s battery indicated 25-50% during a recent interrogation.However, the range was 11-25% previously in the physician's office.The patient's settings were recently lowered, but this was not able to be confirmed as the cause of the fluctuation in the battery status indicator.No further relevant information has been received to date.
|
The data from the two vns tablets involved in the event was reviewed by the manufacturer.The fluctuating battery status indicator was an expected event as the 25% indicator is flagged at 2.87 v for a m106 generator.On 04/19/2018, the battery voltage was at 2.849 v, which is 21 mv below the value at which the 25% indicator is flagged.The diagnostics were within normal limits.Two months later, the battery voltage was at 2.883 v, which is 13 mv above the 25% indicator voltage ceiling.The diagnostics were within normal limits.A rebound of 34 mv within two months was determined to be within reason and is not indicative of a device failure.
|