• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - MARLBOROUGH UPHOLD¿ LITE; MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN Back to Search Results
Model Number M0068318170
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/07/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device has been received for analysis.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that an uphold lite with capio slim was used during an enterocele repair procedure performed on (b)(6) 2018.According to the complainant, during the procedure, upon passing the device through the insertion site without applying force, the suture broke in half.The procedure was completed with another uphold lite with capio slim.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Event Description
It was reported to boston scientific corporation that an uphold lite with capio slim was used during an enterocele repair procedure performed on (b)(6) 2018.According to the complainant, during the procedure, upon passing the device through the insertion site without applying force, the suture broke in half.The procedure was completed with another uphold lite with capio slim.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
A visual analysis of the returned device revealed that the suture on the blue with white stripe dilator was broken, confirming the event.In addition, the remainder of the suture including the dart and the capio slim capturing device were not returned.A review of the device history record (dhr) indicated that the device met all material, assembly, and product specifications at the time of release to distribution.The investigation concluded that this complaint is associated with a product that meets the design and manufacture specifications but due to anatomical/procedural factors encountered during the procedure, performance of the device was limited.The most probable root cause classification is operational context.A labeling review was performed and, from the information available, this device was used per the directions for use (dfu)/product label.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UPHOLD¿ LITE
Type of Device
MESH, SURGICAL, SYNTHETIC, UROGYNECOLOGIC, FOR PELVIC ORGAN PROLAPSE, TRANSVAGIN
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
MDR Report Key7660647
MDR Text Key113149888
Report Number3005099803-2018-02104
Device Sequence Number1
Product Code OTP
UDI-Device Identifier08714729839200
UDI-Public08714729839200
Combination Product (y/n)N
PMA/PMN Number
K122459
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/07/2019
Device Model NumberM0068318170
Device Catalogue Number831-817
Device Lot Number0000047878
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2018
Initial Date Manufacturer Received 06/12/2018
Initial Date FDA Received07/03/2018
Supplement Dates Manufacturer Received08/30/2018
Supplement Dates FDA Received09/20/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age61 YR
-
-