During surgery on an unknown side, the alpha insert popped out of the shell.The surgery was completed with another device.The implantation and explantation dates are left empty as the device involved in this complaint was intra operatively removed.Hence, no expiration date is captured, for the same reason.
|
Correction: date of report, concomitant medical products, date received by mfr, type of report, device evaluated by mfr, additional narrative.Additional: medical products & therapy dates, follow up type.Device history records (dhr): the device manufacturing quality records indicate that the released components met all requirements to perform as intended.No trend considering the following event is identified: assembly issue event summary: it was reported that during surgery, the inlay disenganged from the shell after 5 minutes.Review of received data: no medical data such as x-rays, surgical notes or any other case-relevant documents received.Devices analysis: visual examination: the inlay has been returned for an investigation.It has a deformed peg and some scratches on the outer and inner surface.Additionally, there are scratches on the rim area.On the anchoring surface of the inlay there are only very light imprints of the spikes of the metallic shell.Additionally, these imprints seem to be not in same distance from the polar peg indicating a wrong alignment of the inlay within the shell.See pictures attached.Measurement: to ensure the insert has correct dimensions, relevant characteristic according to the inspection plan was measured with the caliper.Due to the deformation of the peg, further measurements could not be performed.Characteristic feature diameter.Specification: 42.69 +0.05/-0.05 mm.Measured value: 42.69mm.Conclusion: the outer diameter of the insert can be confirmed.Review of product documentation: the compatibility check was performed from www.Productcompatibility.Zimmer.Com and showed that the product combination was approved by zimmer biomet.Surgical technique: liner insertion.The pole plug allows appropriate fitting with the polyethylene liner peg.Bone or soft tissue remnants must not overlap the edge of the titanium shell as they may prevent the insert from snapping into position.The shell edge must be free from any tissue and particular attention must be paid to the posterior inferior bony edge of the acetabulum.Clean and dry the inner surface of the shell, connect the liner to the setting instrument, position the liner over the entrance plane of the shell and rotate clockwise.The peg of the polyethylene liner must be inserted into the pole plug hole.Complete seating of the liner with a light hammer blow.If the liner can still be rotated after light impaction, this indicates mispositionning, nonconcentric, or soft tissues interference between the liner and the shell surfaces.Root cause analysis: root cause determination using rmw: intraoperative complications due to excessive assembly force , leading to wrong aligned inlays.Possible, imprints from the spike indicate a wrong alignment intraoperative complications due to lack of assembly force and/or unclean mating faces are used, leading to wrong aligned inlays or loss of instrument-implant connection.Possible, imprints from the spike indicate a wrong alignment.Moreover these imprint are very light which indicated that the impaction force might have been too low or the mating faces were unclean.Intraoperative complications due to wrong alignment of insert in shell: possible, imprints from the spike indicate a wrong alignment.Intraoperative complications due to wrong positioned insert in shell :possible, imprints from the spike indicate a wrong alignment.Intraoperative complications due to excessive impaction force :possible, excessive force could lead to a failure of the connection.Intraoperative complications due to excessive impaction force :possible, excessive force could lead to a failure of the connection.Intraoperative complications due to lack of adequate impaction force :possible, inadequate force could lead to a failure of the connection.The very light imprints indicate that the impaction force might have been too low.Conclusion summary: it was reported that during surgery, the inlay disenganged from the shell after 5 minutes.Review of the dhr of returned insert with size gg/32 indicates that the devices met all requirements to perform as intended.The performed measurement was found to be according to the specification.Possible causes for the reported failure include high or low load due to impaction during the primary implantation, wrong assembly procedure, slight deformation of the metallic shell due to the very hard bone conditions, soft tissue and/ or debris left between insert and cup prior to seating and storing.The insert in a rather cold place which might lead to a slight decrease of the outer diameter (material reduction).The imprints on the anchoring surface of the inlay coming from the metallic spikes of the shell are very light and asymmetric and the peg ist deformed.Therefore, the most likely root cause of the event is the wrong alignment of the inlay within the metallic shell combined with a too low impaction force.However, ased on the given information and the results of the investigation, an exact root cause could not be determined.The need for corrective measures is not indicated and zimmer gmbh considers this case as closed.Zimmer biomet¿s reference number of this file is (b)(4).
|