• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL OT VERIO IQ METER; GLUCOSE MONITORING SYS/KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL OT VERIO IQ METER; GLUCOSE MONITORING SYS/KIT Back to Search Results
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
On (b)(6) 2018, the reporter contacted lifescan (b)(4), alleging that the subject meter read inaccurately erratic.The reporter claimed obtaining blood glucose readings of ¿425 and 336 mg/dl¿ with the subject meter, performed within 20 minutes of each other.Based on statistical methodology, the calculated difference of these glucose results exceeds lifescan¿s criteria for precision.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no indication that the product caused or contributed to an adverse event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT VERIO IQ METER
Type of Device
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
gubelstrasse 34
zug 6300
SZ  6300
Manufacturer (Section G)
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
gubelstrasse 34
zug 6300
SZ   6300
Manufacturer Contact
mariano chiusano
gubelstrasse 34
zug 6300
SZ   6300
MDR Report Key7662977
MDR Text Key113165816
Report Number3008382007-2018-01904
Device Sequence Number1
Product Code NBW
Combination Product (y/n)Y
PMA/PMN Number
K110637
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Type of Report Initial
Report Date 07/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Other Device ID Number1-1960Z86
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date07/02/2018
Initial Date Manufacturer Received 07/02/2018
Initial Date FDA Received07/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age71 YR
-
-