• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL OT VERIOVUE METER; GLUCOSE MONITORING SYS/KIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL OT VERIOVUE METER; GLUCOSE MONITORING SYS/KIT Back to Search Results
Lot Number 4350151
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
On (b)(6) 2018, the reporter contacted lifescan (b)(4), alleging that the subject meter read inaccurately erratic.The reporter claimed obtaining blood glucose readings of ¿20, 26, 30, 90 and 95 mg/dl¿ with the subject meter, performed within 20 minutes of each other.Based on statistical methodology, the calculated difference of these glucose results exceeds lifescan¿s criteria for precision.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no indication that the product caused or contributed to an adverse event.
 
Manufacturer Narrative
The lay user/patients meter have been returned and evaluated by lifescan product analysis with the following findings: the meter passed all testing with no faults found.The reported issue could not be confirmed.Complaints relating to the reported product(s) were evaluated.It was concluded that the number of complaints for the product(s) did not breach thresholds indicative of a systemic issue.If lifescan obtains additional information regarding this complaint, a follow-up report will be submitted.At this time, lifescan considers this matter closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT VERIOVUE METER
Type of Device
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
gubelstrasse 34
zug 6300
SZ  6300
MDR Report Key7662980
MDR Text Key113168206
Report Number3008382007-2018-01892
Device Sequence Number1
Product Code NBW
Combination Product (y/n)Y
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 06/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Lot Number4350151
Other Device ID Number1-32Y1SY9
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2018
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date06/17/2018
Device Age15 MO
Initial Date Manufacturer Received 06/17/2018
Initial Date FDA Received07/05/2018
Supplement Dates Manufacturer Received06/17/2018
Supplement Dates FDA Received07/31/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-