• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - SAN JOSE GALAXY INTRAVASCULAR ULTRASOUND SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - SAN JOSE GALAXY INTRAVASCULAR ULTRASOUND SYSTEM; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number H749A70200
Device Problem Retraction Problem (1536)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.If there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Event Description
Same case as mdr id#: 2134265-2018-06333 & 2134265-2018-06014.It was reported that automatic pullback failure occurred.An ilab ultrasound imaging system was used in conjunction with an imaging catheter and a pullback sled to view the target lesion.During the procedure, the imaging catheter was able to produce image.However, overload message was displayed during pullback.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAXY INTRAVASCULAR ULTRASOUND SYSTEM
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - SAN JOSE
150 baytech drive
san jose CA 95134
Manufacturer (Section G)
BOSTON SCIENTIFIC - SAN JOSE
150 baytech drive
san jose CA 95134
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7663896
MDR Text Key113323377
Report Number2134265-2018-06334
Device Sequence Number1
Product Code IYO
Combination Product (y/n)N
PMA/PMN Number
K980851
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 06/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH749A70200
Device Catalogue NumberA7020
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/18/2018
Initial Date FDA Received07/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-