STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL13.7
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Lens was returned in liquid, in lens vial.Visual inspection found the optic torn and portion of the lens missing.One similar complaint type event was reported for units within the same lot.(b)(4).
 
Event Description
The reporter indicated that a 13.7mm micl13.7, -13.0 diopter, implantable collamer lens tore during loading.There was no patient contact and the backup lens was implanted with no issues.Cause of event was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
MDR Report Key7663994
MDR Text Key113234669
Report Number2023826-2018-01005
Device Sequence Number1
Product Code MTA
UDI-Device Identifier00841542103626
UDI-Public00841542103626
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 06/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date07/31/2019
Device Model NumberMICL13.7
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/02/2018
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/07/2018
Initial Date FDA Received07/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-