Same case as mdr id: 2134265-2018-06499.It was reported that the speed was unstable.A rotalink¿ advancer and a rotablator console were selected for use.During preparation, the cables were connected respectively to the console and the compressed air pressure is evaluated at 100 psi.Then, the physician introduced the rotawire to the proximal end of the 1.50mm rotalink¿ burr and advancer already mounted obtaining it in its back.The draw test then was realized; however, it was noted that the burr rotates in fluctuating revolutions.The set operational speed was at 150,000 rpms but it went from 220,000 to 120,000 in less than 3 seconds with the pedal fully depressed.Also, the sound of revolutions was unusual from the advancer.It was assessed that everything was well connected with the required pressure so the physician tried again obtaining the same results.Physician decided to remove the rotaburr and disconnected the advancer from the rotalink burr.It was then noted that the rotalink burr was slightly bent in a segment proximal the handshake connection.The procedure was completed with a 1.75mm rotalink¿ plus.No patient complications were reported and patient's status was stable.
|