Model Number 106
Device Problem Energy Output To Patient Tissue Incorrect (1209)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/13/2018
Event Type  malfunction  
Event Description
A physician reported that, upon running diagnostics on a patient's generator, she had seen "low output current." impedance was within normal limits and the patient's output current was 1.75 ma.After the diagnostics were ran again, everything checked out normally.No further relevant information has been received to date.
 
Event Description
Data from the physician's tablet including the patient's most recent office visit was received and reviewed.Diagnostics performed on the device all showed ok communication, ok output status, ok impedance, and ok battery status."low" output current was not observed from the data.No additional relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 106
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key7664379
MDR Text Key113247808
Report Number1644487-2018-01145
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750061
UDI-Public05425025750061
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/06/2019
Device Model Number106
Device Lot Number5308
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 06/13/2018
Initial Date FDA Received07/05/2018
Supplement Dates Manufacturer Received07/27/2018
Supplement Dates FDA Received08/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age27 YR
-
-