Model Number S7
Device Problem Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/08/2018
Event Type  malfunction  
Manufacturer Narrative
No parts have been received by the manufacturer for evaluation.
 
Event Description
Medtronic received information regarding a navigation device being used for a sacroiliac and thoracolumbar procedure.The issue occurred intraoperatively during the select patient/acquire scans task and delayed the surgery by less than one hour.It was reported that a spin was taken but it was not transferring to the navigation device.The device was sitting on ¿waiting to acquire¿.The system was rebooted, and the spin was transferred.The surgery was completed with navigation and there was no impact on patient outcome.
 
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.Testing revealed that the reported issue could not be replicated.The system then passed the system checkout and was found to be fully functional.
 
Manufacturer Narrative
The logs for the navigation system were reviewed by medtronic personnel.However, the logs provided no additional insight into the probable cause of the anomaly.The behavior of the reported issue was found to be consistent with a known anomaly in the medtronic navigation software anomaly tracking database.This issue was documented in a medtronic navigation software anomaly tracking database.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7664921
MDR Text Key113283880
Report Number1723170-2018-03173
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 08/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/08/2018
Initial Date FDA Received07/05/2018
Supplement Dates Manufacturer Received07/05/2018
08/13/2018
Supplement Dates FDA Received07/23/2018
08/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/16/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age17 YR
Patient Weight56
-
-