• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS; PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS; PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VICH12.6
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/31/2018
Event Type  malfunction  
Manufacturer Narrative
This product is manufactured in the u.S.But not marketed in the u.S.(b)(4).
 
Event Description
The reporter indicated that, while implanting a 12.6mm vich12.6 implantable collamer lens, diopter -9.0 into the patient's right eye (od), the lens tore/broke.This occurred on (b)(6) 2018.The lens was removed intraoperatively.On (b)(6) 2018 and alternate lens was implanted.Reportedly, no patient injury occurred.The problem is resolved.
 
Manufacturer Narrative
Device evaluation: the lens was returned dry in a micro-centrifuge vial.There was clear surgical residue on the product and lens surface.Visual inspection found the haptic torn.Method code corrected: lens work order search: no similar complaint type events reported for units within the same lot.(b)(4).
 
Manufacturer Narrative
(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer (Section G)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
cynthia thai
1911 walker avenue
monrovia, CA 91016
8002927902
MDR Report Key7665042
MDR Text Key113283664
Report Number2023826-2018-01014
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeTU
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date07/31/2020
Device Model NumberVICH12.6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/06/2018
Initial Date FDA Received07/05/2018
Supplement Dates Manufacturer Received07/23/2018
08/29/2018
Supplement Dates FDA Received08/20/2018
09/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/15/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age23 YR
-
-