MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP MMT-1712B; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-1712B
Device Problems Break (1069); Component Missing (2306); Scratched Material (3020)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/30/2017
Event Type  malfunction  
Manufacturer Narrative
Unit received with minor scratched lcd window, scratched display window cover, keypad overlay texture damage, faded serial number label, missing retainer, missing reservoir tube o-ring, cracked reservoir tube lip and scratched case.Unable to test for displacement test or lock reservoir in place due to missing retainer.The insulin pump involved in this event is the paradigm real-time 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
Customer reported via phone call physical damage to the insulin pump.The blood glucose level was unknown.It was reported the connection with the reservoir compartment was broken, the ring was missing and the display was scratched.The insulin pump will need to be replaced.The device will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
640G INSULIN PUMP MMT-1712B
Type of Device
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
8185464805
MDR Report Key7672186
MDR Text Key113659983
Report Number3004209178-2018-89409
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign
Reporter Occupation Other
Type of Report Initial
Report Date 07/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1712B
Device Catalogue NumberMMT-1712B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2017
Initial Date Manufacturer Received 08/30/2017
Initial Date FDA Received07/09/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-