• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION ADVANCED PERFUSION SYSTEM 1; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1 Back to Search Results
Model Number 801764
Device Problems Failure to Run on Battery (1466); Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/16/2018
Event Type  malfunction  
Manufacturer Narrative
Per the manufacturer's subsidiary, the battery icon was red.
 
Event Description
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the unit was not switching over to battery when there was loss of power.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
 
Manufacturer Narrative
As per manufacturer's subsidiary, there was a loose connection on the power manager board.The connection was tightened and the unit operated as intended.
 
Manufacturer Narrative
The reported complaint was confirmed.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
MDR Report Key7673233
MDR Text Key113517917
Report Number1828100-2018-00363
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K022947
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 09/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number801764
Device Catalogue Number801764
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/18/2018
Initial Date FDA Received07/10/2018
Supplement Dates Manufacturer Received07/09/2018
09/04/2018
Supplement Dates FDA Received08/03/2018
09/17/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-