• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. CARELINK; PROGRAMMER, PACEMAKER Back to Search Results
Model Number 2090AA4
Device Problems Computer Software Problem (1112); Device Displays Incorrect Message (2591); Operating System Becomes Nonfunctional (2996)
Patient Problem No Patient Involvement (2645)
Event Date 06/14/2018
Event Type  malfunction  
Manufacturer Narrative
Product analysis: analysis was able to confirm the reported programmer crashing and displaying error messages.The hard drive was replaced.The device passed final functional and system tests.No other anomalies were found.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the programmer crashed while attempting to interrogate an implantable pulse generator (ipg) and then displayed an error message.The programmer was rebooted and then showed an additional error message.The programmer was returned for service.There was no patient involvement.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Type of Device
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7674421
MDR Text Key113750307
Report Number2182208-2018-01262
Device Sequence Number1
Product Code KRG
UDI-Device Identifier00721902824692
UDI-Public00721902824692
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number2090AA4
Device Catalogue Number2090AA4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2018
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/14/2018
Initial Date FDA Received07/10/2018
Supplement Dates Manufacturer Received08/08/2018
Supplement Dates FDA Received10/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/24/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-