• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP SHOULDER; GLENOID, HEAD W/RETAINING SCREW, RSP, 36MM, -4MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP SHOULDER; GLENOID, HEAD W/RETAINING SCREW, RSP, 36MM, -4MM Back to Search Results
Model Number 508-36-103
Device Problems Unstable (1667); Compatibility Problem (2960)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/18/2018
Event Type  Injury  
Event Description
Revision surgery - due to the patient being unstable.The components were changed to a different sizing.
 
Manufacturer Narrative
The reason for this revision surgery was due to the patient becoming unstable.The previous surgery and the revision detailed in this investigation occurred 32 days apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to the patient becoming unstable.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Type of Device
GLENOID, HEAD W/RETAINING SCREW, RSP, 36MM, -4MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, TX 78758-5445
MDR Report Key7675469
MDR Text Key113591933
Report Number1644408-2018-00665
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888912024778
UDI-Public(01)00888912024778
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/21/2023
Device Model Number508-36-103
Device Catalogue Number508-36-103
Device Lot Number871C1165
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 06/18/2018
Initial Date FDA Received07/10/2018
Supplement Dates Manufacturer Received07/17/2018
Supplement Dates FDA Received07/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/25/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
508-00-436, LOT 919F1905
Patient Outcome(s) Other; Required Intervention;
Patient Age72 YR
-
-