Device Problem Infusion or Flow Problem (2964)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 06/12/2018
Event Type  malfunction  
Manufacturer Narrative
A gehc service representative performed a checkout of the equipment and confirmed the reported complaint.The ez change manifold was replaced.No report of patient involvement.The initial reporter is located outside the u.S., and therefore this information is not provided due to country privacy laws.
 
Event Description
The hospital reported that higher than expected level of carbon dioxide was noted.There was no report of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AVANCE
Type of Device
ANESTHESIA GAS MACHINE
Manufacturer (Section D)
GE HEALTHCARE
3000 n grandview blvd.
waukesha WI 53188
Manufacturer (Section G)
GE HEALTHCARE
3000 n grandview blvd.
waukesha WI 53188
Manufacturer Contact
john szalinski
3000 n grandview blvd.
waukesha, WI 53188
8479711808
MDR Report Key7677201
MDR Text Key113640853
Report Number2112667-2018-01356
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K032803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 07/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 06/12/2018
Initial Date FDA Received07/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/29/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-